Valhalla scientific |
|
Àå ºñ ¸í
|
AC/DC Current Calibrator
|
¸ð
µ¨ ¸í
|
2555A
|
Á¦
Á¶ »ç
|
Valhalla scientific
|
|
|
|
|
|
Àåºñ »ç¾ç |
|
|
Delivers 1mA to 100 amps DC or RMS AC
Specifications: Dynamic Range: 0 to 200% of range up to maximum rated output Maximum Output Current: 100A DC or RMS AC DC Accuracy: (180 days 25°C ?5°C) all ranges <100A ?.015% of range ?.03% of output 100 Amp Range: ?.03% of range ?.03% of output AC Accuracy: (180 days)( 25°C ?5°C) ?.15% of range ?0.1% of output to 100Hz ?.2% of range ?0.2% of output to 400Hz ?.3% of output ?0.3% of range to 1KHz Input Impedance: 100 Kilohms Input/Output Ratio: 2.00000 volt input provides full scale output (1.0000 volt input produces 100.0 amps output) Stability: ?.0005% - 15 min. (?.005% for 24 Hours) Maximum Input: 3 VDC or 2 VAC (RMS) Response Time: 1 millisecond to ?.01% of final value following input amplitude or frequency change Maximum Isolation Voltage: ?200 VDC or Peak AC Input Common Mode Rejection Ratio: 60db @ DC linearly decreasing to 40 db @ maximum frequency Load Regulation: Output current changes less than ?.002% Temperature Range: 0°C to 50°C Temperature Coefficient: ?.001% of output ?.002% of range/°C below 20°C and above 30°C (double above for temperature coefficient for AC) Compliance Voltage: 4 Volts Size: 266mm/10?" L x 432mm/ 17" W x 584mm/23" D Weight: 45 Kg/100 lbs. Net 66 Kg/140 lbs. shipping wt. Power: 115/230 VAC ?10% 50-60Hz 2000 watts
|
|
|
|
|